HygroGen2

Portable humidity and temperature generator to calibrate humidity and temperature measuring instruments (multi-point calibration)

Download Datasheet »

Contact me about this product Request Product Demonstration

Portable humidity and temperature generator to calibrate humidity and temperature measuring instruments (multi-point calibration)

Applications
The new HygroGen2 is a portable generator for calibration of humidity and temperature measuring instruments. The device sets a new standard in portable calibration. The HydroGen2 works like a "mobile calibration laboratory" and is intended for companies that regularly need to calibrate a large number of probes. The calibrator allows simply, fl exible calibration with the advantage that the calibrated units can be integrated in the operative processes again quickly. The HygroGen offers numerous outstanding benefi ts particularly to the pharmaceutical industry.

Features
• Generates a stable reference environment Range of application: 0…60 °C, accuracy: ±0.05 K
• Reaches equilibrium humidity in typically 5 minutes
• Suitable for all humidity and temperature probes
• Calibrates up to 5 probes simultaneously
• Easy-to-use trouchscreen monitor
• DVI interface for external monitor
• USB interface for connection of keyboard, mouse and ROTRONIC USB probes
• The integrated HW4 software ensures easy calibration and adjustment of all probes.
• External heated connections for a dew point mirror reference are standard. This allows the user to adjust the reference probes with extremely high precision or to reduce the total calibration uncertainty.
• The water quality is kept at a high level by a UV sterilizer, meaning algae and bacteria cannot form.

Optional FEATURES

AutoCal: 
Automated calibration of up to 6 HC2 probes connected via AC3001 USB adaptor: 
- Option to automatically adjust 1 temperature and 10 humidity points 
- Creates PDF calibration certificate for each probe under test 
- 20 user programs (up to 200 set points per program)

Range Extension
- Temperature range extension -5...60 °C (Standard 0...60 °C) 
- Humidity range extension 2...99 %RH (Standard 5...95 %RH)

   

Control range max.

5…95 %rh and 0…60 °C

Control range min.

5…95 %rh over the range of 10…50 °C

Control stability

<0.02°C, 0.1%rh

Temperature gradient

<0.05 (15...50 °C), <0.1 (5...60 °C), 0.15 (0 °C)

Humidity gradient

0.1 %rh or better

Typical time to set-point

5 minutes (35 to 80 %rh change, 23 °C) 10 minutes (23 to 45 °C change)

Temperature sensor

PT100 1/3 DIN Class B

Humidity sensor

Hygromer

Control probe accuracy

±0.8 %rh(10…95 %rh) / ±0.1 °C at 23°C

Desiccant

Molecular sieve

Saturator

Front panel fill, on-screen level indication

Water quality

Distilled or deionised water

Water volume

approx. 80 ml

Chamber volume

ca. 2 Liter

External interface

USB (9 ports), Ethernet (service use only)

Power supply / Power consumption

110...230 VAC, 50...60 Hz, 3 A

External connections

6mm temperature controlled sample loop connections

Enclosure material

Powder coated aluminium

Dimensions

450 x 406 x 205mm

Weight

13.0000